Laboratory Liquid Handling Equipment Servicing – Lab Innovations UK

Laboratory Liquid Handling Equipment Servicing

Description​

Microlab Servicing

To maintain and prolong the life of your laboratory instruments, it is important to have them serviced on a regular basis.

Microlab Technologies have been at the forefront of providing a full calibration and service of Hamilton’s
Microlab instruments since 2003. Our engineers are fully trained by and supported by Hamilton Company.

Test Equipment:

• We use Pipette Wizard software in conjunction wirth an Ohaus Explorer precision balance for the calibration and
validation of the instrument. The calibrations are conducted as an ‘as found’ test in accordance with ISO 8655-6
over 3 test volumes at 10, 50 and 100% aliquots of the total syringe volume to identify errors and confirm linearity, precision
and accuracy of the dispensing functions according to the manufacturers agreed metrics for inaccuracy and imprecision.

• The Ohaus balance is validated with the use of a 1gm and 5gm weight, which are annually validated and certified by the
NMO (National Measurement office)

• Testing is conducted using DI water and is monitored using a Testo immersion probe and humidity probe for temperature and
humidity, a Druck pressure meter is used to register the barometric pressure of the environment in mbars.

• The equipment is validated annually and certificates are provided as part of the service and calibration report.